Technical and functional
descriptions
|
| Technical features |
MK-1
|
MK-1-1
|
MK-1-2
|
| tidal volume (VT), ml |
50–1200
|
50–1200
|
50–1200
|
| respiration rate (f), min-1 |
6-30
|
6-60
|
6-60
|
| minute ventilation volume
(V), l/min |
0.3–25
|
0.3–25
|
0.3–25
|
| inhalation/exhalation ratio
(ÒI:ÒÅ) |
1:1, 1:2, 1:3
|
4:1-1:4
|
4:1-1:4
|
| max. airway pressure, cmÍ2Î |
50
|
5-50
|
5-50
|
| inspiration pause, % |
–
|
0-50
|
0-50
|
| flow trigger, l/min |
–
|
2-20
|
2-20
|
| PEEP, ñmÍ2Î |
0-15
|
0-30
|
0-30
|
| inhalation pressure limitation,
cmÍ2Î |
50
|
5-50
|
5-50
|
| emergency oxygen supply, l/min |
40-70
|
40-70
|
40-70
|
|
|
Monitoring
|
| Graphical loop V/P |
–
|
+
|
+
|
| Graphical loop _/V |
–
|
+
|
+
|
| Graphical airway pressure
waveform |
–
|
+
|
+
|
| Graphical flow waveform |
–
|
+
|
+
|
| Electronic numerical pressure
data |
–
|
+
|
+
|
| Electronic volume measure
|
–
|
+
|
+
|
| ÑÎ2
concentration waveform |
–
|
±*
|
+
|
| Pleth. waveform |
–
|
+
|
+
|
| Tidal volume |
+
|
+
|
+
|
| Minute ventilation volume |
–
|
+
|
+
|
| Respiration rate |
+
|
+
|
+
|
| Airway pressure (Ppeak, Pmean,
Pplateau, PEEP) |
–
|
+
|
+
|
| Compliance |
–
|
+
|
+
|
| Resistance |
–
|
+
|
+
|
| Inspiration attempt indication |
–
|
+
|
+
|
| Inhalation/exhalation ratio
(ÒI:ÒÅ) |
+
|
+
|
+
|
| Inspiration Î2
concentration |
–
|
+
|
+
|
| Inhalation/exhalation ÑÎ2
and N2O concentration |
–
|
±*
|
+
|
| Anesthetic vapors minimal
alveolar concentration (ÌÀC) |
–
|
±*
|
+
|
|
|
Alarms
|
| High and low airway pressure |
–
|
+
|
+
|
| High and low minute ventilation
volume |
–
|
+
|
+
|
| High and low VÒ |
–
|
+
|
+
|
| High and low respiration rate
f |
–
|
+
|
+
|
| Low Î2
concentration |
–
|
+
|
+
|
| High and low ÑÎ2,
N2O and anesthetic vapors
concentration |
–
|
±*
|
+
|
| Low Î2
input pressure |
+
|
+
|
+
|
| Electric drive malfunction
|
+
|
+
|
+
|
| Respiratory circuit depressurization
|
–
|
+
|
+
|
| APNEA |
–
|
+
|
+
|
|
* - optional
|
|
|
|
Êëàññè÷åñêèé
äûõàòåëüíûé êîíòóð
|
"Ïëîñêèé"
äûõàòåëüíûé êîíòóð
|
|
Ventilation modes
HR, ECG ( I, II, III, aVR, aVL, aVF, V1-V6), arrhythmia analysis
and ST-segment level detection, SpO2, Respiration rate, NIBP, Temperature
°C, audio and visible alarms.
|
HR, ECG ( I, II, III, aVR, aVL, aVF, V1-V6),
arrhythmia analysis and ST-segment level detection, SpO2,
Respiration rate, NIBP, Temperature °C, audio and visible
alarms.
|
|

|
|