VIAN artificial lung ventilation machines

Technical and functional descriptions
Technical features
MK-1
MK-1-1
MK-1-2
tidal volume (VT), ml
50–1200
50–1200
50–1200
respiration rate (f), min-1
6-30
6-60
6-60
minute ventilation volume (V), l/min
0.3–25
0.3–25
0.3–25
inhalation/exhalation ratio (ÒI:ÒÅ)
1:1, 1:2, 1:3
4:1-1:4
4:1-1:4
max. airway pressure, cmÍ2Î
50
5-50
5-50
inspiration pause, %
0-50
0-50
flow trigger, l/min
2-20
2-20
PEEP, ñmÍ2Î
0-15
0-30
0-30
inhalation pressure limitation, cmÍ2Î
50
5-50
5-50
emergency oxygen supply, l/min
40-70
40-70
40-70
Monitoring
Graphical loop V/P
+
+
Graphical loop _/V
+
+
Graphical airway pressure waveform
+
+
Graphical flow waveform
+
+
Electronic numerical pressure data
+
+
Electronic volume measure
+
+
ÑÎ2 concentration waveform
±*
+
Pleth. waveform
+
+
Tidal volume
+
+
+
Minute ventilation volume
+
+
Respiration rate
+
+
+
Airway pressure (Ppeak, Pmean, Pplateau, PEEP)
+
+
Compliance
+
+
Resistance
+
+
Inspiration attempt indication
+
+
Inhalation/exhalation ratio (ÒI:ÒÅ)
+
+
+
Inspiration Î2 concentration
+
+
Inhalation/exhalation ÑÎ2 and N2O concentration
±*
+
Anesthetic vapors minimal alveolar concentration (ÌÀC)
±*
+
Alarms
High and low airway pressure
+
+
High and low minute ventilation volume
+
+
High and low VÒ
+
+
High and low respiration rate f
+
+
Low Î2 concentration
+
+
High and low ÑÎ2, N2O and anesthetic vapors concentration
±*
+
Low Î2 input pressure
+
+
+
Electric drive malfunction
+
+
+
Respiratory circuit depressurization
+
+
APNEA
+
+

* - optional

Êëàññè÷åñêèé
äûõàòåëüíûé êîíòóð
"Ïëîñêèé"
äûõàòåëüíûé êîíòóð

 

 

The MK anesthesia machines are intended to conduct inhalation anesthesia with the use of oxygen, nitric oxide and anesthetic vapors both for adult and pediatric patients.


Ventilation modes

HR, ECG ( I, II, III, aVR, aVL, aVF, V1-V6), arrhythmia analysis and ST-segment level detection, SpO2, Respiration rate, NIBP, Temperature °C, audio and visible alarms.

 

 

 

 

 

Patient monitor UÌ-300:
HR, ECG ( I, II, III, aVR, aVL, aVF, V1-V6), arrhythmia analysis and ST-segment level detection, SpO2, Respiration rate, NIBP, Temperature °C, audio and visible alarms.